BPC300-C Photochromic Contact Lens Spectrophotometer
Measurement Specifications
Spectral range of operation | 350-780nm, 280nm-780nm optional, contact us for other ranges |
Bandwidth | 5nm |
Wavelength accuracy | ± 0.15nm |
Beam diameter at sample plane | 5mm |
Measured transmission range | 0.5-100% |
Conditioning beam | AM2 with requirements of ISO8980-3:2013 0-50klx Shutter Optional filtering |
Temperature range | 20-50°C |
Temperature stability | ± 0.2°C |
Bench space required | 1m deep x 1.5m wide for bench top assembly, see Dimensions for details |
Services requirements | 5 x main sockets 1200W power consumption Dry gas supply where measurements below ambient are required |
Monochromatic Probe Source
Probe source | 250W Quartz Halogen lamp, Deuterium source optional for 280nm |
Monochromator configuration | Symmetric, single Czerny-Turner source, second matched monochromator for acquisition |
Monochromator focal length | 300mm |
Diffraction gratings | 1800 g/mm standard, additional options available |
Diffraction grating mount | Triple grating turret |
Dispersion | 4.05nm/mm |
Bandwidth | 5nm standard, additional options available |
Wavelength accuracy | ± 0.15nm standard |
Chopping frequency | 175 Hz (typical) |
Beam diameter at sample plan | 5mm diameter |
Water Bath
Temperature range | 20-50°C |
Temperature stability | ± 0.2°C |
Temperature accuracy | ±1°C calibratable and supplied with calibrated thermometer and mount |
Conditioning Beam
Spectral match | AM2 with requirements of ISO8980-3:2013 |
Beam illuminance | 0-50klx |
Beam filtering | Optional filtering |
Number of configurable apertures | 4 |
Number of configurable filters | 6 |