Electro-optical

Spectral range of operation

300-780 nm

Spectral data interval

5 nm

Spectral bandwidth (FWHM)

5 nm

Wavelength accuracy

± 0.3 nm

Wavelength reproducibility

± 0.05 nm

Stray light rejection at 2.5 FWHM

10-4

Photo-detectors

Multi-alkali PMT and Silicon photodiode

Photo-current measurement

Dual-channel six decade picoammeter

Measured quantity

Spectral radiance

Minimum measurement distance

200 mm

Maximum measurement distance

>10 m

Fields of view

11, 5 and 1.7 mrad

Range of blue light radiance (typ.)

10-100000 W.m-2.sr-1

System requirements

Bench space

0.5x 0.5 m not including working space for computer and luminaire under test

Computer

OS: Windows 7 or newer (32-/64-bit)

Minimum hard disk space: approx. 100MB

Minimum RAM: 2GB

1 x USB 2.0 ports

Power

2 x 110/220V AC mains sockets, 400VA total

Mechanical & Interface

Dimensions

 

Software control

Benwin+ spectral acquisition software

PSL Wizard measurement guidance, calculation and reporting application

Interface

USB 2.0

Share this: