Optical Layout

Configuration

Pair symmetrical Czerny Turner monochromators, additive dispersion

Focal length

- Total focal length: 300mm

- Individual units: 150mm

Grating mount

Kinematic

Number of gratings

1 pair

Grating size

33x33mm

Aperture ratio

f/4

Number of entrance/exit ports

1 entrance, 1-2 exits

Spectral range

Within range 200nm-2400nm

Opto-Mechanical

Wavelength scanning

By integrated stepping motor and micro-stepping drive electronics

Slit type

Fixed, micrometer or motorised variable

Slit width & height

10µm-10mm (W) x 20mm (H)

Filter size

- 6 position filter wheel : 25mm diameter

- 8 position filter wheel : 22mm diameter

Optical Performance (quoted for typical 2400 g/mm gratings)

Spectral range

200-600nm

Linear dispersion

1.35nm/mm

Wavelength accuracy

± 0.15nm, ± 0.05nm with software correction

Wavelength reproducibility

± 0.05nm

Resolution full/reduced slit height

0.25/0.1nm

Stray light rejection at 2.5 FWHM

10-8

Wavelength sweep speed

30nm/s

487 Dual-Channel Picoammeter Module - Electrical

Channel 1 input

Current input to picoammeter

Channel 2 input

Current input to picoammeter

Gain ranges

1010-105 V/A

Maximum current Input

100µA

Frequency response

DC to 30Hz

Gain accuracy

+1%

Gain stability

200ppm/°C

Output stability

5ppm/°C to 500ppm/°C depending on sensitivity

Linearity

< 0.025% departure from linearity from zero to full scale

ADC resolution

4½ digit BCD (0 to 19999) i.e. > 14 bit resolution

ADC integration time

100ms

Auxiliary input range

0- 9.8V

Picoammeter input impedance

Virtual earth

Auxiliary input impedance

1 MΩ

415 High Voltage Power Supply Module - Electrical

Output voltage range

-350 to -1500V

Maximum current

1.5mA

Temperature stability

100 ppm/°C

Stability to voltage fluctuations

25ppm for 10% change in line voltage

Ripple and noise

100ppm peak-peak