LuxCal250 Lux and Luminance Meter Calibration
LuxCal250
| Lamp type | Quartz Tungsten Halogen lamp, G6.35 base |
| Nominal lamp power and voltage | 250W, 12V |
| Operating current | 10.4A (adjusted to obtain 2856K operation) |
| CCT | 2856±20K |
| Expected lifetime | 2000 hours |
| Calibration frequency | 100 hours use/ one year recommended |
| Dimensions | 600L x 500W x 700H (mm) |
| Typical illuminance level one (no filter) | 19000 lx |
| Typical illuminance level two (ND1) | 1800 lx |
| Typical illuminance level three (ND2) | 180 lx |
| Typical illuminance level four (ND3) | 20 lx |
| Typical illuminance level five (ND4) | 2 lx |
LumiCal50
| Lamp type | Grit-blasted Quartz Tungsten Halogen lamp, G6.35 base |
| Nominal lamp power and voltage | 50W, 12V |
| Operating current | 4.0A (adjusted to obtain 2856K operation) |
| CCT | 2856±20K |
| Expected lifetime | 2000 hours |
| Calibration frequency | 100 hours use/ one year recommended |
| Typical luminance level one (no filter) | 21000 cd.m-2 |
| Typical luminance level two (ND1) | 1800 cd.m-2 |
| Typical luminance level three (ND2) | 200 cd.m-2 |
| Typical luminance level four (ND3) | 30 cd.m-2 |
| Typical luminance level five (ND4) | 3 cd.m-2 |
LuxMeter+
| Spectral function | V(λ) photometric |
| V(λ) matching error | f1 ≤ 3 % |
| Typical responsivity | 0.5 nA/lx |
| Diffuser diameter | 7mm |
| Cosine matching error | f2 ≤ 1.5 % |
| Typical minimum detectable illuminance | 20 mlx |
| Typical maximum detectable illuminance | 20 klx |
| Display | 5 digits, auto-ranging |
| Power supply | Mains transformer |
| Interface | USB 2.0 |
610 Constant Current Power Supply
| Input voltage | 100V - 220V, 5A max, 50-60Hz |
| Output voltage & current | 0-10.4A in remote use. Max Voltage: 26V |
| Connection type | IEC mains input connection |
| Resolution | 0.1A local; 0.001A over SCPI protocol |
| Operation modes | Remote mode PC: SCPI protocol allows the Set Current to be assigned to an accuracy of 1mA Local mode Pre-set: Allows pre-set Current settings to be selected Variable: Used to set the current between 0-10.4A, indicated on the display |
| Current pre-set | 4.00 A, 5.40 A, 6.30 A, 8.50 A, 10.40 A, 0-10.4A variable |
| Max output power | 250W |
| Setting control | Remote setting by user interface controlled through USB Local setting by rotary switch (fixed current or variable settings) |
| Operating conditions | Ambient temperature: 0°C - +40°C Storage temperature: -20°C - +85°C |
| Drift | 0.05% of rated output (Over an 8-hour interval with constant line, load and temperature, after a 30-minute warm-up period) |
| Temperature stability | 30ppm/°C |
| Compliance to standards | EN 61010-1:2010 EN 61326-1:2013 |
| Accuracy | ±(0.04% +0.05A) |
| Output power ripple | <0.1% |
| Dimensions | 450L x 300W x 130H (mm) |
| Weight | 5 kg |
| Mode types | Constant Current, Constant Voltage |