ORM400 Picoammeter
Electrical
Channel 1 input | Current input to picoammeter |
Channel 2 input | Current input to picoammeter |
Auxiliary input | Voltage input to ADC |
Gain ranges | 1010-105 V/A |
Maximum current Input | 100µA |
Frequency response | DC to 30Hz |
Gain accuracy | +1% |
Gain stability | 200ppm/°C |
Output stability | 5ppm/°C to 500ppm/°C depending on sensitivity |
Linearity | < 0.025% departure from linearity from zero to full scale |
ADC resolution | 4½ digit BCD (0 to 19999) i.e. > 14 bit resolution |
ADC integration time | 100ms |
Auxiliary input range | 0- 9.8V |
Picoammeter input impedance | Virtual earth |
Auxiliary input impedance | 1 MΩ |