Ge-CAL Calibrated Germanium Reference Standard

Ge-CAL in PVE systemThe Ge-CAL is a general purpose germanium photodiode offering excellent linearity and low noise, calibrated in spectral response (A.W-1) over the wavelength range of 800-1800nm. Calibration is performed with respect to a detector of the same type calibrated by Physikalisch-Technische Bundesanstalt (PTB).

Its' mount is compatible with the entire range of Bentham monochromators and accessories.

This detector is recommended for use with the 400 Series of detection electronics.

Specfication

PHOTODIODE
Active area 5mm diameter
Max. Operating Temperature -40 to +70°C
Temperature dependence of responsivity Up to 3%/ °C at band edge
Window material Glass
TYPICAL VALUES
Peak Wavelength 1540nm
Peak Responsivity 0.9 A.W-1
Dark Current <1pA
Max Shunt Resistance 15 kΩ
NEP 1.5 x 10-12 W/Hz1/2
CALIBRATION
Measurement Type Spectral responsivity under short-circuit conditions
Wavelength Range 800-1800nm, 5nm intervals
Calibration Frequency 3-yearly recommended
Temperature at Calibration 20°C

Spectrum

Si-CAL spectra