Get Adobe Reader


The DH-PbSe-Te detector is a large area (3 x 3mm) thermo-electrcially cooled detector. It is partuclarly suitable for accurate spectroradiometry and general spectrometry applications over the 1µm to 5µm range.

This device requires a.c. coupled detection electronics, typically comprising optical chopper and lock-in such as the Bentham 218 and 225, and a bias power supply such as the 215.

Sandwich detectors or 'two-colour' detectors are also available. An infrared transmitting silicon detector is mounted in front of the PbSe detector extending the wavelength covered by a single package down to less than 250nm

 

Specification
Wavelength Peak l p
4.3-4.5 µm
D* (l p , 750, 1)
≥7.00E9 cmHz 1/2W-1
Blackbody D* (500K, 750, 1)
≥9.3E8 cmHz 1/2W-1
Responsivity @l p
≥4000 V/W
Resistance
0.2-5.0 MW
Time Constant
3-10 µsec
Operating Temperature
253 K
Recommended Range
1µm - 5µm
Cooler controller type CPS1 or CPS1M also required