Migration from legacy SDK

This page is for customers with existing code written against the legacy Bentham SDK - the pyBen Python module, the bendll.h/dllerror.h/dlltoken.h C headers, or the BI_*.vi LabVIEW wrapper VIs - who want to move to this SDK.

The guide covers the languages that both SDKs target: Python, C, C++ and LabVIEW.

What does not change

Both SDKs are wrappers around the same BI_* functions exported by the BenHW DLL, so the hardware side of your application carries over unchanged:

  • Your system configuration (.cfg) and setup (.atr) files are used as they are.

  • The call sequence is the same: build the system model, load the setup, initialise, park, zero calibrate, select wavelength, measure, close. See SDK Model.

  • Every token and error code defined in the legacy dlltoken.h/dllerror.h exists in this SDK with the same name and the same numeric value. See Tokens and Errors.

  • Component IDs passed to get/set and other functions are the IDs in your .cfg file, as before.

What changes is how you load the library, how results come back, and how errors are reported.


Python: pyBen to benhw

Installation

Uninstall the pyBen wheel or MSI installer and install benhw from PyPI. benhw requires Python 3.12 or later and bundles the BenHW DLLs, so nothing needs to be copied onto PATH.

pip uninstall pyBen
pip install benhw

Overview of the differences

Topic

pyBen

benhw

Import

import pyBen

from benhw import BenHW, tokens, SdkException

Calling a function

Module-level function: pyBen.initialise()

Method on an instance: hw = BenHW() then hw.initialise()

Function names

BI_ prefix removed

Unchanged from pyBen: BI_ prefix removed

Return value

A dict: {"Error": 0, "reading": 1.23}

The value itself (1.23), a result object for several outputs, or None

Errors

Check result["Error"] != 0 after every call

An exception derived from SdkException is raised

Tokens

Raw integers, e.g. 11

Named constants, e.g. tokens.MonochromatorCurrentWL

DLL loading

On import; DLL must be found by the module

On BenHW(); uses the bundled DLL, or BenHW(dll_path=...)

Before and after

Legacy pyBen code:

import pyBen

r = pyBen.build_system_model("system.cfg")
if r["Error"] != 0:
    raise RuntimeError(r["description"])
pyBen.load_setup("system.atr")
pyBen.initialise()
pyBen.park()
pyBen.zero_calibration(300.0, 800.0)

pyBen.select_wavelength(555.0, 0)
result = pyBen.automeasure()
if result["Error"] == 0:
    print(result["reading"])

wl = pyBen.get("mono", 11, 0)["value"]      # 11 = MonochromatorCurrentWL
pyBen.close()

The same code using benhw:

import time
from benhw import BenHW, tokens, SdkException

hw = BenHW()
try:
    hw.build_system_model("system.cfg")
    hw.load_setup("system.atr")
    hw.initialise()
    hw.park()
    hw.zero_calibration(300.0, 800.0)

    settle_ms = hw.select_wavelength(555.0)
    time.sleep(settle_ms / 1000)
    reading = hw.automeasure()
    print(reading)

    wl = hw.get("mono", tokens.MonochromatorCurrentWL, 0)
except SdkException as e:
    print(f"BenHW error {e.error_code}: {e.message}")
finally:
    hw.close()

Error handling

Remove the ["Error"] checks. Every method either succeeds or raises:

  • SdkException is the base class; it has error_code (the BI_* value) and message.

  • Each error code has its own subclass, e.g. AdcOverloadException for BI_ADC_Overload or InvalidComponentException for BI_invalid_component. These are importable from benhw or benhw.exceptions. See Errors for the full mapping.

  • When the DLL returns the generic BI_error (-1), the wrapper calls BI_report_error() for you and raises the more specific exception if one is available. The legacy pattern of calling report_error() after a failure is no longer needed.

  • build_system_model no longer returns the description string; if the configuration cannot be built, the DLL’s message is the message of the raised SdkException.

from benhw import exceptions

try:
    reading = hw.automeasure()
except exceptions.AdcOverloadException:
    hw.autorange()
    reading = hw.automeasure()

Functions whose signature or result changed

Most functions only need the pyBen. prefix replaced and the ["key"] lookup removed. The following need more attention:

pyBen

benhw

select_wavelength(wl, delay)

select_wavelength(wl). The delay argument is gone. The settle delay in milliseconds is now returned (pyBen discarded it); wait for it before measuring.

multi_select_wavelength(wl, delay)

multi_select_wavelength(wl), returning the settle delay as above.

component_select_wl(id, wl, delay)

component_select_wl(id, wl).

get_zero_calibration_info() returns {"calibrationInfo": {"wavelength": (...), "darkCurrent": (...), "ADCOffset": (...)}}

Returns a result object with list attributes wavelength, DarkCurrent and ADCOffset. Note the capital D in DarkCurrent. multi_get_zero_calibration_info(group) is the same.

multi_measurement() returns a tuple under "reading"

Returns a list[float], one reading per group.

read(size, id) returns "message"

read(buffer_size, id) returns a result object with buffer (bytes) and chars_read (int).

trace(on)

trace(i, LoggingDir): also takes the directory the trace log is written to.

version() returns {"description": ...}

Returns a str.

report_error()

Returns the detailed error code as an int. Usually unnecessary (see above).

get/set with integer tokens

Integers still work, but use the names in benhw.tokens for readability.

benhw also exposes functions that pyBen never wrapped, including get_str/set_str, multi_automeasure, multi_park, get_min_step, get_max_bw, logging (start_log, stop_log, get_log) and SCPI_query/SCPI_write. See Python API for the complete list.


C: bendll.h to benhw.h

The function names and arguments are the BI_* functions you already call, so most C code compiles after changing the includes and the loader calls. Install the headers and DLLs as described in C API.

Overview of the differences

Topic

Legacy headers

benhw.h

Includes

bendll.h, dllerror.h, dlltoken.h

benhw.h (includes errors.h and tokens.h)

Loading the DLL

Call init() and check for LOADED_DLL

Automatic on the first BI_* call, or benhw_load(NULL) to load up front

Unloading the DLL

close()

benhw_unload()

DLL used

benhw32_stdcall.dll (WINAPI)

benhw64.dll or benhw32_cdecl.dll, chosen by process bitness

Settle delay type

long*

int*

BI_read sizes

short BufferSize, short* CharsRead

uint16_t buffer_size, uint16_t* chars_read

Function not found in DLL

init() fails with UNABLE_TO_LOAD_FUNCTIONS

That call returns BENHW_ERR_DLL_LOAD_FAILED (-9000)

Before and after

Legacy code:

#include <windows.h>
#include "dllerror.h"
#include "dlltoken.h"
#include "bendll.h"

int main(void)
{
    char error[256];
    long settle_delay = 0;
    double reading;

    if (init() != LOADED_DLL)
        return 1;

    if (BI_build_system_model("system.cfg", error) != BI_OK) { /* ... */ }
    BI_load_setup("system.atr");
    BI_initialise();
    BI_park();

    BI_select_wavelength(555.0, &settle_delay);
    Sleep(settle_delay);
    BI_automeasure(&reading);

    close();        /* legacy close() only unloads the DLL */
    return 0;
}

The same code using benhw.h:

#include <stdio.h>
#include "benhw.h"

int main(void)
{
    char error[256];
    int settle_delay = 0;
    double reading;

    if (!benhw_load(NULL))          /* optional: fail early if the DLL is missing */
        return 1;

    if (BI_build_system_model("system.cfg", error) != BI_OK)
    {
        printf("%s\n", error);
        return 1;
    }
    BI_load_setup("system.atr");
    BI_initialise();
    BI_park();

    BI_select_wavelength(555.0, &settle_delay);
    Sleep(settle_delay);
    BI_automeasure(&reading);

    BI_close();     /* release the system model */
    benhw_unload();
    return 0;
}

Notes

  • Remove the legacy loader helpers. Delete calls to init() and close(): they were helper functions defined in bendll.h, not DLL exports. The legacy close() only called FreeLibrary; it never called BI_close(). Call BI_close() yourself before unloading.

  • Change the settle-delay variable type. Use int instead of long for BI_select_wavelength, BI_multi_select_wavelength and BI_component_select_wl.

  • Ship the cdecl or 64-bit DLL, not benhw32_stdcall.dll. benhw.h looks for the DLL next to your executable, then in a dlls subfolder, then on the normal DLL search path. To use a specific file, call benhw_load("C:\\path\\to\\benhw64.dll") before any other call. Your program can now be built as 64-bit.

  • Error and token constants (BI_OK, BI_error, MonochromatorCurrentWL, …) keep their names and values. Error checks such as if (result != BI_OK) are unchanged.

  • If another part of your process has already loaded the DLL, use benhw_load_from_module(hmodule) instead of benhw_load().


C++: bendll.h to benhw.hpp

C++ code that used the legacy C headers can follow the C migration above and keep calling BI_* functions. Alternatively, switch to the benhw::BenHW class in benhw.hpp (in the same package), which replaces error-code checks and output buffers with return values and exceptions:

#include <iostream>
#include <thread>
#include <chrono>
#include "benhw.hpp"

int main()
{
    benhw::BenHW hw; // loads the DLL; unloads it when hw goes out of scope
    try {
        hw.build_system_model("system.cfg");
        hw.load_setup("system.atr");
        hw.initialise();
        hw.park();
        hw.zero_calibration(300.0, 800.0);

        int settle_ms = hw.select_wavelength(555.0);
        std::this_thread::sleep_for(std::chrono::milliseconds(settle_ms));
        double reading = hw.automeasure();
        std::cout << reading << "\n";

        double wl = hw.get("mono", MonochromatorCurrentWL, 0); // tokens.h constant

        hw.close();
    } catch (const benhw::SdkException& e) {
        std::cerr << "BenHW error " << e.error_code() << ": " << e.what() << "\n";
        return 1;
    }
    return 0;
}

The method names are the C function names without the BI_ prefix. The exception classes and result types (e.g. get_zero_calibration_info_result) match those of the Python SDK. See C++ API.


LabVIEW: BI_*.vi wrappers to the .NET assembly

The legacy LabVIEW examples were VIs (BI_SystemSetup.vi, BI_MoveToWavelength.vi, BI_TakeReading.vi, BI_Get.vi, BI_Set.vi, …) that called the DLL through Call Library Function Nodes and returned integer error codes. This SDK replaces them with the .NET assembly BenHW.dll, which LabVIEW calls through .NET Constructor and Invoke Nodes. Install it and create the constructor as described in LabVIEW.

Overview of the differences

Topic

Legacy VIs

.NET assembly

Calling a function

Call Library Function Node on the BenHW DLL

Invoke Node on the reference from the BenHW Constructor Node

Function names

BI_select_wavelength

PascalCase, without BI_: SelectWavelength

Outputs

Pointer parameters wired to indicators

Return value of the Invoke Node

Errors

Integer return value compared with BI_OK

SdkException reported in the node’s error out cluster

DLL selection

Path configured in each Call Library Function Node

Single dllPath input on the Constructor Node

Common replacements

Legacy VI / DLL call

.NET method (Invoke Node)

BI_build_system_model, BI_load_setup, BI_initialise (BI_SystemSetup.vi)

BuildSystemModel, LoadSetup, Initialise

BI_park (BI_Park.vi)

Park

BI_select_wavelength (BI_MoveToWavelength.vi)

SelectWavelength: returns the settle delay in milliseconds

BI_automeasure (BI_TakeReading.vi)

Automeasure: returns the reading

BI_get / BI_set (BI_Get.vi / BI_Set.vi)

Get / Set, with the same numeric token values

BI_close_shutter (CloseShutter.vi)

CloseShutter

BI_report_error (BI_Return_Error.vi)

Not needed: the error is in error out. ReportError is still available.

BI_version (BI_GetVersion.vi)

Version: returns a string

BI_close (BI_Close.vi)

Close, then close the .NET reference

A scan (BI_Scan.vi) becomes a loop of SelectWavelength, a wait for the returned settle delay, and Automeasure. The full method list is in .NET API.

Notes

  • Errors arrive in the error cluster. When a method throws, LabVIEW reports a .NET exception in the Invoke Node’s error out; the exception message identifies the BenHW error. Wire error in/error out through your sequence, so a failure stops the following calls, in place of the legacy checks on each return code.

  • Match the DLL to LabVIEW’s bitness. Wire the path of benhw64.dll to dllPath for 64-bit LabVIEW, or benhw32_cdecl.dll for 32-bit LabVIEW. Both are in the downloaded zip.

  • Requires the .NET Framework 4.8 runtime.

  • Tokens are entered as numbers, as before. Values are unchanged; see Tokens.