474 Transformer & Amplifier Module

  • 474 modular transformer and amplifier unit
  • 474 transformer unit

Given a signal having both AC and DC components, the 474 transformer allows de-coupling of the two, with the DC being blocked. Gain can then be applied to the sought after AC signal to ensure optimal recovery.

Voltage biasing of the detector used can be achieved by applying voltage across the primary coil of the transformer.

Core benefits

Features

  • Free-standing transformer unit
  • Single width module housed within the 417 unit
  • Signal gain provided by transformer and low noise amplifier
  • Decouples DC bias source generated photocurrent
  • Signal gain provided by transformer and low noise amplifier
  • Near-short circuit device operation

Electrical

Amplifier voltage Gain

500

Amplifier bandwidth (-3dB)

5Hz to >100kHz

Amplifier short circuit input noise

<1nV Hz-1/2 at 1kHz

Amplifier maximum output

10V

Transformer DC resistance

0.05Ω

Transformer gain

~100

Optimal transformer frequency

600Hz

Interface & Mechanical

Dimensions

Single width module, 3U high

Connector

BNC

Ordering Information

S400_474

474 transformer

S400_474_PREAMP

474 fixed gain low noise voltage amplifier module